Version: 10.08.01 ======== Contact: joerg.dubbert@physik.uni-muenchen.de ======== Documentation for the LMM Local Database ========================================= - The Local Template version 0.0 of 01.01.2001 is used - The original Tube table has been deleted - Contains the LocBat, QcOpPar, PlugBat, Tube and WirBat tables from the JNR local database - The Tube table from the JNR database is renamed to TubeJNR_local The RepMeasMarker is set if the tube has been remeasured at LMM - LMM measurements are contained in the new TubeLMM_local table It contains the following fields: IdTube tube id RejCode rejection code, see RejCode table QcDateTime date/time of Qc QcOp operator QcWdyA wire displacement y-coord. side A at 0 degree meas. QcWdzA wire displacement z-coord. side A at 0 degree meas. QcWdyB wire displacement y-coord. side B at 0 degree meas. QcWdzB wire displacement z-coord. side B at 0 degree meas. QcWdyA180 wire displacement y-coord. side A at 180 degree meas. QcWdzA180 wire displacement z-coord. side A at 180 degree meas. QcWdyB180 wire displacement y-coord. side B at 180 degree meas. QcWdzB180 wire displacement z-coord. side B at 180 degree meas. QcWirFrq uncorrected wire frequency QcTemp temperature during wire frequency meas. QcGasLeak gas leakage (currently not filled) QcCurrLeak leakage current at 3400 V in nA QcCosmic rate in cosmic ray test (not filled) Comment tube location, see Comment table - New CommentLMM table LMM tube kept at LMM MPI tube send to MPI RIP tube destroyed at LMM - New OcOpLMM table contains the abrevations and names of the QC operators - MPI measurements are contained in the new TubeMPI_local table It contains the following fields IdTube tube id IdLayer layer id AsTubeLoc tube location QcDateTime date/time of Qc QcQcWirFrq uncorrected wire frequency QcTemp temperature during wire frequency meas. AsWirYSiteRO y wire position measured during assembly AsWirYSiteHV y wire position measured during assembly